Input/Output Interconnections

All input/output signals are accessible at J001
accessory connector. With the provided P001 mating
connector
make
2
AWG 24 (0.2 mm
) - or smaller wire - to the pins
indicated in the following figure to obtain the desired
capability.
Input signals
Pin 3-7
Remote high voltage ON/OFF or high
voltage interlock; requires a permanent
closed pin contact.
Pin 3-7 must be shorted to allow the high voltage
output if no remote/interlock contact is connected.
Output signals
Pin 1-6
Current recorder output 1 (pin 1 signal out,
pin 6 ground). The output voltage 0 to + 10
Vdc is proportional to the pump current 0 to
10 mA (e.g. 10 Vdc correspond to 10 mA).
The
resolution
(corresponding to 10 µA).

Pump Connection

The pump is connected to the controller rear panel via
the high voltage connector by a coaxial high voltage
cable assembly.
the
connections
NOTE
is
10
mV
minimum
P001 Mating Connector and J001 Accessory Connector
The front panel REC.OUT connector is a duplicate of
this signal on a BNC connector.
with
Pin 2-6
Current recorder output 2 (pin 2 signal out,
pin 6 ground). The output voltage 0 to + 10
Vdc is proportional to the pump current 0 to
1 mA (e.g. 10 Vdc correspond to 1 mA). The
resolution
10 mV minimum (corresponding to 1
Pin 4-8
Voltage recorder output (pin 4 signal out, pin
8 ground). The output voltage 0 to 5 V is
proportional to the controller high voltage
output 0 to 5 kV (e.g. 1 V corresponds to 1
kV in linear mode).
Pin 5-9
High voltage indication contact.
contact (which carries 1 A at 250 Vac and
0.2 A at 30 Vdc) is open when the high
voltage is off and closes when the high
voltage in on.
High voltage can cause severe injury or death.
Before installing or removing the connector:
- Turn the power off.
- When connecting the cable to 2 l/s pumps, or
larger, be sure that the ground spring is in place on
the high voltage connection of the pump.
- Install the pump connector and secure
it with screws.
- Plug the controller connector.
22/24
TECHNICAL INFORMATION