Table 4 Guidance and manufacturer's declaration – electromagnetic immunity –
for ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacture's declaration – electromagnetic immunity
The CHECKY is intended for use in the electromagnetic environment specified below.The
customer of the user of the CHECKY should assure that it is used in such an environment .
Immunity test
IEC 60601
test level
Conducted RF
3 Vrms
150 kHz to
IEC 61000-4-6
80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to
2.5 GHz
Compliance
Electromagnetic environment -
level
guidance
Portable and mobile RF communications
equipment should be used no closer to
any part of the CHECKY, including
cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of
the transmitter.
Recommended separation distance
3 Vrms
d = 1.167
3 V/m
d = 1.167
d = 2.333
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacture
and d is the recommended separatio n
distance in metres (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey, should be
less than the compliance level in each
frequency range.
Interference may occur in the vicinity of
equipment marked with the following
symbol:
123
80 MHz to 800 MHz
800 MHz to 2.5 GHz