TURBO-V 2K-G OUTLINE

The following figure shows the Turbo-V 2K-G outlines (dimensions are in mm [inches]).

INTERCONNECTIONS

The following paragraphs describe the Turbo-V
2K-G interconnections.
J1 ñ REMOTE I/O
At power-on the controller goes into ìRemote I/Oî
mode, even if it was in ìSerialî mode before
power-down.
This connector carries all the input and output sig-
nals to remote control the Turbo-V 2K-G.
It is a 15-pins D type connector; the available sig-
nals are detailed in the table, the following para-
graphs describe the signal characteristics and use.
PIN N.
SIGNAL NAME
1
START/STOP (+)
2
START/STOP (-)
3
INTERLOCK (+)
4
INTERLOCK (-)
5
SPEED SETTING (+)
6
FAULT COMMOMN
7
SOFT START (+)
8
SET POINT
9
+24 Vdc
10
PURGE (+)
11
SET POINT
12
VENT (+)
13
FAULT N.O.
14
ANALOG OUTPUT
15
GROUND
73
TECHNICAL INFORMATION
INPUT/
ISOLATION
OUTPUT
IN
Optocoupled
IN
Optocoupled
IN
Optocoupled
IN
Optocoupled
IN
Optocoupled
OUT
Relay
IN
Optocoupled
OUT
Relay
OUT
Transformer
IN
Optocoupled
OUT
Relay
IN
Optocoupled
OUT
Relay
OUT
Optocoupled
OUT
Transformer
87-900-968-01(C)