Controller-to-Pump Connection

To connect the controller to the pump an appropriate
interconnection cable must be prepared to connect the
J9 terminal board of the controller with the pump
connector.
The following figure shows the pump output connector
configuration where pins:
Pins A-F = pump temperature sensor
Pins B-C-D = 54 Vac 3-phase output to pump motor
stator
Pin E = ground.
Controller-to-pump cable
54
TECHNICAL INFORMATION
87-900-871-01 (B)