Calibration - LDS Ebulliometer Notice D'utilisation

Ebulliomètre électrique
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

3 - Calibration

The calibration allows to measure the boiling point of a standard wine to calibrate the disk or the calcula-
tion software and initialize the atmospheric pressure sensor. This operation must be repeated each time the
device is turned on.
• Rinsing
- Check that the drain valve is correctly closed.
- Refill the boiler with the standard wine to the bottom of the condenser.
Then empty the boiler using the drain valve.
• Boiling point of the standard wine
- Refill the boiler to the bottom of the meniscus above the red line.
- Turn on the instrument using the on/off switch.
The main screen appears.
- Let cold water circulate into the condenser, with a regular flow, not
excessive.
- Press the heating button during
The following screen appears successively :
Verif ier chaudiere
1111
The heating (shown by the symbol ) lasts about 7 minutes, during which
the temperature rise can be seen on the screen.
To stop heating during the operation, press again on the heating button.
- When the boiling temperature is stable, the device makes a beep to
indicate the end of the measurement. Heating stops automatically.
- The boiling temperature T
- Match the alcohol content of the standard wine with its boiling temperature
T
on the disc, or use the programme «EBULLIOLOG», entering the
Standard
values «Reference Temperature» and «Alcoholic reference».
- Drain.
- Press briefly the heating button to come back to the main screen. The
device is ready for a sample measurement.
• Example
Standard wine with official alcohol degree = 9,2 % vol (± 0,05% vol.)
T
= 93,1°C
Standard
Calibrate the disc as indicated in the picture opposite.
Model 160350T with traditional thermometer :
Press the heating button 5 seconds to launch heating. When the wine is boiling, check on the thermometer
that the temperature is stable (after 1 to 2 minutes). Once it is stable, record the boiling temperature T
Turn off the heating using the heating switch. Adjust the disk or the programme as indicated above.
Repeat calibration every 2 hours.
Notice Ebulliometer - Version 2012 - Réf. 160350T/160350D - Page 14
at least 8 seconds
CALIBRATION
6666
remains on the display.
Standard
CALIBRATION
93,10 C
to launch calibration.
Never turn on the
instrument
the heating switch
when the boiler is empty or
when draining. You might
damage
the
instrument
if you do not observe this
rule.
For more accuracy,
i
we advise you to turn
on the instrument
about 15 minutes
before use.
i
The accuracy can
be reached with
standard
whose
properties
are similar to the wine to be
tested. If no standard wine is
available, it can be replaced
by
demineralized
déminéralisée, considered as
standard wine with alcoholic
content of 0% but accuracy
will be lower.
using
wine,
water
.
Standard

Publicité

Table des Matières
loading

Table des Matières