Setting Up Of The Instrument - LDS Ebulliometer Notice D'utilisation

Ebulliomètre électrique
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Instrument :
1. Introduction of samples
2. Condenser upper outlet
3. Condenser
4. Cold water inlet
5. According to the model:
Mercury thermometer
Temperature probe
6. Thermometer stopper
7. Boiler
8. Heating cartridge
9. Drain valve
10. LCD screen for temperature
(Model Ref.160350D)
11. Heating switch
12. 6-plug DIN connector – connection of the
temperature probe (Model Ref.160350D)
13. Adapter connector - Power supply 24V DCa
14. ON/OFF switch

2 - Setting up of the instrument

• Installation of the instrument
- Put the instrument on a flat surface
- Tightly and carefully push in the boiler on the cartridge holder. You may add
some water on the washers to make this operation easier.
- Connect the instrument on the power supply.
- Brancher l'appareil sur secteur.
• Installation of the probe
Place the probe on the boiler without modifying the position of the silicon
gasket on the probe, adding some water on the silicon gasket if necessary in
order to push it in at the maximum. The end of the probe must be situated
between 0,5 and 1 cm above the end of the heating element.
In case of the model with mercury thermometer, place it the same way,
making sure that the mercury bulb or the probe is 1 cm above the heating
cartridge.
In both case, the measuring stem must be on the same axis than the heating
cartridge.
• Installation of the cooling system and tubing
Attach the water connections to the condenser.
 Cold water inlet to the bottom;
 Water outlet to the top.
Attach additional tubing to the boiler drain and place the tubing into a sink to
drain easier.
Accessories:
15. Disc for electric ebulliometer
16. Standard wine for calibration
Réf. 160350/1
17. Defoamer for ebulliometer
18. Chromosulfuric acid
Réf. 690117E
19. Power cable
Réf. 160350/2
20. Cooling system and drain hoses
Réf. 160350/3
21. Dropping-pipette with bulb
Réf. 160350/1
Réf. 160350/4
Notice Ebulliometer - Version 2012 - Réf. 160350T/160350D - Page 13
(Supplied conditionally)
Réf. 160350/5
Réf. 100550
Réf. 990001
Réf. 972500
Réf. 630011
Installation of boiler on
stand
Do not remove
the probe or the
thermometer
between a calibration
and a measurement.