Annexe - Endress+Hauser Liquisys M CLM223 Manuel De Mise En Service

Transmetteur de conductivité
Masquer les pouces Voir aussi pour Liquisys M CLM223:
Table des Matières

Publicité

Annexe

Function group
CALIBRATION
C
MEAS. VALUE DISPLAY
with
TEMPERATURE DISPLAY
in °C
Function group
SETUP 1
A
Function group
SETUP 2
B
Function group
CURRENT INPUT
Z
Z
Function group
CURRENT OUTPUT
O
Function group
ALARM
F
Function group
CHECK
P
118
13
Annexe
Calibration temperature
Calibration
entry (MTC)
InstF
25.0 C
°
= Installation factor
C1 (3)
–35.0 ... +250.0 C
°
C131
Calibration temperature
entry
Cellc
(if B1 = fixed)
= Cell constant
C1 (2)
.
25.0 C
°
–35.0 ... +250.0 C
°
C121
Residual coupling
Start calibration
AirS = Airset
C1 (1)
Current meas. value
C111
Temperature display
in °F
1st error is displayed
(if present)
Selection of operation mode
Selection of unit
displayed
c
ond = conductive
ind = inductive
ppm
; mg/l; %; TDS; none
MOhm = resistance
(% only if A1 = conc)
conc = concentration
A1
Selection of temperature
Selection of temperature
measurement
.
Pt100
Pt1k (= Pt 1000)
NTC30 (= NTC 30 k
)
fixed
B1
Cont. switch-off
Delay for cont.
by current input
switch-off current input
Off;
Input
0 s
0 ... 2000 s
Z1
Current output selection
Select measured variable
for 2nd current output
Out1
; Out2
°C; mS/cm; Contr
O2
O1
Select contact type
Select alarm
delay unit
Stead = steady contact;
Fleet = fleeting contact
s
; min
F1
Switch polarisation
Set alarm threshold
detection on or off
off
; on
Off;
Low: High; Lo+Hi;
P1
Lo!; Hi!; LoHi!
Entry of
α
value of
Entry of correct
conductivity value of
calibration solution
calibration solution
.
2.10 %/K
Current meas. value C133
0.00 ... 20.00 %/K
0.0 µS/cm ... 9999 mS/cm
C132
Entry of correct
Entry of a
α
value of
calibration solution
.
2.10 %/K
0.00 ... 20.00 %/K
C122
Display of residual coupling
Calibration status
(Airset)
is displayed
.
o.k.
0.0 µS
C112
E---
Temperature display
Measured value display
suppressed
Current output in %
Other errors are displayed
(up to 10 errors)
Selection of unit
Display format selection
displayed
(if A1 = conc)
auto;
µS/cm; mS/cm; S/cm;
XX.xx
; X.xxx; XXX.x; XXXX
µS/m; mS/m; S/m
auto ; k
×
cm; M
(omitted if A1 = conc)
A2
A3
Entry of a value
Entry of correct
(if B2 = linear)
process temperature
(if B1 = fixed)
2.10 %/K
0.00 ... 20.00 %/K
25.0
°C
–35.0 °C ... +250.0 °C
B3
B2
Delay for cont.
Switch-off limit value
switch-on current input
for current input
0 s
0 ... 2000 s
50%
0 ... 100%
Z2
Z3
Table option selection
Characteristic selection
read
edit
table
O3 (3)
Simulation value entry
sim = simulation
O3 (2)
current value
0 ... 22.00 mA
Current range selection
lin = linear
O3 (1)
4–20 mA;
0–20 mA
Alarm delay
Error current
setting
0 s (min)
0 s ... 2000 s (min)
22 mA
F2
(depends on F2)
F3
2.4 mA
Enter alarm delay
Set lower
alarm threshold
0 s (min)
0 µS/cm
0 ... 2000 s (min)
P3
0 ... 9999 mS/cm
P2
Liquisys M CLM223/253
Display of calculated
Calibration status
installation factor
is displayed
1.0
o.k.;
0.10 ... 5.0
E---
C134
Calibration status
Display of calculated
is displayed
cell constant
o.k.;
0.0025 ... 99.99 1/cm
C123
E---
C124
Store calibration
results
yes;
no; new
C113
C114
Uncompensated
Measured value display
measured value
is displayed
Current output in mA
Entry of cell constant
Entry of cable
resistance
cond / ind / MOhm
(if A1 = cond)
1.000 / 1.98 / 0.01 1/cm
0.0025 ... 99.99 1/cm for
0.00
×
cm; k ×
m
cond; ind; MOhm
0.00
... 99.99
A5
A4
Temperature sensor
Enter temperature
calibration (omitted
difference (omitted
if B1 = fixed)
if B1 = fixed)
Display of actual value
Current offset
–35.0 ... +250.0
°C
–5.0 ... 5.0 °C
B4
B5
Switch-off direction
Feedforward control
for current input
to PID controller
Low;
High
Off;
lin = linear
Z4
Z5
Entry of number of
Selection of
value pair in table
value pairs in table
1
1
1 ... 10
1 ... number of value pairs
asign
O331
O332
O321
Entry of 0/4 mA value
Entry of 20 mA value
0 µS/cm / 0 k ·cm /
2000 mS/cm / 500 k ·cm /
0 % / 0 °C
9999 % / 150.0 °C
entire measuring range
entire measuring range
O311
O312
Error number
Set alarm contact
selection
to be effective
1
yes
; no
F4
1 ... 255
F5
Set upper
Select process
alarm threshold
monitoring
9999 µS/cm
Off;
AC; CC; AC+CC
0 ... 9999 mS/cm
P4
P5
AC!; CC!; ACCC!
C135
C125
A6
B6
Z6
O333
O313
F6
P6
A0027482-FR