Diagnostics - Troubleshooting - Breakdowns - Tecno-gaz PURA Mode D'emploi

Masquer les pouces Voir aussi pour PURA:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

ENGLISH
DIAGNOSTICS – TROUBLESHOOTING - BREAKDOWNS
TROUBLESHOOTING TABLE
ATTENTION: WITH AUTOCLAVE SWITCH SWITCHED ON
Anomaly
st
1
Autoclave is
not requesting
water
nd
2
Supply tap
The
closed
device
produces
no water
rd
3
No power
st
1
Low
waterworks
The
pressure
device
nd
2
Defective or
produces
spent osmotic
very little
membrane
water
The examples mentioned above are typical anomalies that may occur. Everything
outside the third column pertains to Tecno-Gaz S.p.a.
Cause
(possibly adopted by the user)
Check that the max level led of the
autoclave is off (see PAR 8).
Check that the tap supplying the
device is not closed by mistake.
Check that you have inserted the
connector correctly into the
autoclave
Check that the waterworks
pressure is not below 2 bar.
NA
Contact the Service centre.
Normal solution
16
Solution by TECNO-
GAZ authorized
technicians
-
-
Check that the 230
VAC on the contact
1 e 2, in the
autoclave, is
present.
Find the cause of
the low pressure
and eliminate it.
Replace the osmotic
membrane.

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

O1001npu

Table des Matières