Symptom
Fiber bundle will not
install in Light Source
Audible "clicking" or
arcing sound during
the lamp ignition
process

Lamp Replacement

Note
1.
Turn off power, unplug from power source, and wait several minutes to cool.
Not a compatible fiber bundle
Install/Remove Multi Link™
Expansion Port
Multi Link™ Expansion Port
Installed
No problem
WARNING During lamp ignition, 25KV is present across the electrodes.
WARNING Always wear eye protection when removing, replacing, or handling
the lamp.
WARNING DO NOT TOUCH OR TAMPER WITH THE LAMP. The CL Series
lamps are highly pressurized, any damage or improper handling may cause
premature failure or explosion of the lamp.
WARNING Dropping or mishandling of the CL Series lamps during replacement
will break the lamp. HANDLE WITH EXTREME CARE.
WARNING The CL Series lamps are hot immediately after use. Allow the lamp
to cool for five (5) minutes after operation. Temperatures peak in the CL Series
light sources approximately 60 seconds after the Light Source is powered off.
WARNING CL Series Replacement Lamps are not interchangeable and use
the correct Lamp for the Light Source wattage.
Disconnect from the supply mains before removal of the CL Series lamp.
Use only CL 100 and CL 300 Replacement Lamps.
A standard slot head screwdriver is required to change the CL Series lamps.
To get maximum optical performance, the fixture must be fully seated and the
captive screw 1/4 turn past finger tight.
The CL Series lamps are designed to assure correct orientation and cannot be
installed incorrectly.
Cause
See Operating Instructions
Use compatible bundle design
See Operating Instructions
Remove Multi Link™ Expansion Port
Normal operation
CL100 / CL300 Surgical Illuminator
Remedy