Hach 8310 Manuel De L'utilisateur page 3

Masquer les pouces Voir aussi pour 8310:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 36
Table of Contents
1 Specifications
on page 3
2 Expanded manual version
3 General information
on page 4
4 Transmitter
on page 5
Section 1 Specifications
Specifications are subject to change without notice.
Applications
-1
K (cm
)
Accuracy
Transmitter measurement range
Pt100 temperature response (t
90%)
Maximum temperature (°C)
Maximum pressure (bars)
Sample connection
Applications
-1
K (cm
)
Accuracy
Transmitter measurement range
Pt100 temperature response (t
90%)
Maximum temperature (°C)
Maximum pressure (bars)
Sample connection
Material
Maximum temperature (°C)
Maximum pressure (bars)
on page 4
Table 1 Probe specifications
8310 / 8315
Pure and ultra pure water
0.01 to 200 μS.cm
< 30 seconds
125 (8310)
150 (8315)
10 (8310)
25 (8315)
¾ inch NPT
8312 / 8317
Waste water and drinking water
1 μS to 20 mS.cm
< 3 minutes
125 (8312)
150 (8317)
10 (8312)
25 (8317)
¾ inch NPT
Table 2 Flow chamber specifications
5 Installation and startup
6 Maintenance
on page 11
7 Spare parts
on page 12
0.01
< 2%
-1
Food and pharmaceutical industries
1.0
< 2%
-1
08313=A=0001
PVC
60 at 2 bar
15 at 25°C
on page 5
8311 / 8316
Moderately conductive solutions
0.1
< 2%
0.1 μS to 2 mS.cm
< 45 seconds
125 (8311)
150 (8316)
10 (8311)
25 (8316)
¾ inch NPT
8394
(sterilized)
0.01
< 2%
-1
0.01 to 200 μS.cm
< 45 seconds
150
25
Tri-Clamp 1½ or 2 inch
08318=A=0001
316 L SS
150
25
English 3
-1

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

831583118316831283178394

Table des Matières