The electrode exposed to the air reaches an equilibrium corresponding to the partial pressure of oxygen
and thus to saturation in water at the given temperature. The instrument shows the measured saturation,
current, temperature, and will calibrate automatically when readings are stable.
3.
After rinsing the electrodes with distilled water, immerse them in the samples and read the display.
Stirring the solution during the measurements promotes the homogeneity and is obligatory! The
advection rate must be at least 10 cm/s.
4.
Rinse the electrode always after use and store it in distilled water.

Good Laboratory Practice

1.
Press CALIBRATE (F1) to go to the calibration menu
2.
Select GLP and press OK or SELECT (F2) to display a complete calibration report.

Data Storage in the internal memory

Storing data
1.
Press STORE to store the displayed value in memory. The display shows a next storage address.
2.
Select the desired address and press OK or SELECT (F2) to enter the measured value in memory.
Recalling data
Press  to recall the stored data. The display shows a list of stored values.
1.
2.
Select the desired address to read the stored values.
3.
Press OK to return to the normal measurements.