Error Indication - Gram C6 Série Manuel D'utilisation

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

3
Setup F.S.
Zero point calibration:
4
Press
light is on
5
F.S. calibration: (Press
directly if the value to be
6
input is same as F.S. when
stable light is on)
7

ERROR INDICATION

1
Err 1
2
Err 2
3
Err 3
4
Err 5
5
bAt-Lo
6
Err 8
7
Err 10
when the stable
Load standard weights.
Calibration finished.
AD value is too small when calibration.
Zero point is out of range when calibration.
Zero point is out of range when power on
Weight input is 0 when full scale calibration
Indicator out of battery, please charge!
Mistakes in inputting F.S.
Resolving capability is too high
Input F.S.:Press
the digit bit; press
[fULL ]
the digit; press
the input F.S. e.g.: 30000
[noLoAd] Assure there's no loading
Input weight of standard weights
loaded: Press
the digit bit; press
the digit. Until the input value is
[AdLoAd]
same as the loaded weight and
the digit bit is at the highest bit,
press
stable light is on.
to choose
to select
to confirm
to choose
to select
to confirm when the
11