Guidance and manufacturer's declaration

Guidance and manufacturer's declaration – electromagnetic emissions
The Human Care lift is intended for use in the electromagnetic environment specified below. The customer or the user of the
Human Care lift should assure that it is used in such an environment.
Emission test
RF emissions
CISPR 11
RF emissions CISPR 11
Harmonic emissions IEC 61000-
3-2
Voltage fluctuations/flicker
emissions IEC 61000-3-3
Guidance and manufacturer's declaration – electromagnetic immunity
The Human Care lift is intended for use in the electromagnetic environment specified below. The customer or the user of the
Human Care lift should assure that it is used in such an environment.
Immunity test
IEc 60601 test level
Electrostatic
+/- 6 kV contact
discharge (ESD)
+/- 8 kV air
IEC 61000-4-2
Electrical fast
+/- 2 kV for power
transient / Burst
supply lines
IEC 61000-4-4
+/- 1 kV for input/
output lines
Surge
+/- 1 kV differential
mode
IEC 61000-4-5
+/- 2 kV common
mode
Voltage
<5 % UT
dips, short
(>95 % dip in UT)
interruptions
for 0,5 cycle
and voltage
variations on
40 % UT
power supply
input lines
(60 % dip in UT)
IEC 61000-4-11
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT (>95 % dip
in UT))
for 5 sec
www.humancaregroup.com
compliance
Group 1
Class B
Class A
Complies
compliance level
+/- 6 kV contact
+/- 8 kV air
+/- 2 kV for power
supply lines
n/a. for input/
output lines
+/- 1 kV differential
mode
n/a. for common
mode
<5 % UT
(>95 % dip in UT)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT))
for 5 sec
Electro -
magnetic environment - guidance
The Human Care lift uses RF energy only for its RF
emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
The Human Care lift is suitable for use in all
establishments including domestic establishments
and those directly connected to the public low-voltage
power supply network that supplies buildings used for
domestic purposes.
Electro -
magnetic environment - guidance
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material,
the relative humidity should be at least 30 %.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment. If the user of the
Human Care lift requires continued operation during
power mains interruptions, it is recommended that the
lift be powered from an uninterruptible power supply or
battery.