Read the displayed value. A resistance < 120
measuring continuity.
5.4 Diode Measurement
Insert the black test lead into the negative COM terminal and the red
test lead into the positive V terminal.
Set the function switch to
Select the diode mode with the MODE button.
Connect the test leads with the diode to be measured. The forward
voltage will indicate 0.4V ~ 0.7V. Reverse voltage will indicate "OL".
Shorted devices will indicate near 0mV and an open device will indicate
"OL" in both polarities.
red
probe
forward test

5.5 Data Hold Function

Press HOLD to freeze the value onto the display. Press again to resume
normal operation.
5.6 MAX Hold
Press MAX to freeze the highest value onto the display. Press again to
resume normal operation.

5.7 Manual Ranging

Press the RANGE button to set the range manually. Press and hold the
button to return to the auto ranging mode.
5.8 Backlight
Press and hold the
5.9 Battery Replacement
Remove the screws using an appropriate Philips head screwdriver.
Replace the batteries with 2 x AAA batteries.
Reassemble the meter before use.