Hach LANGE sc200 4-20 Manuel D'utilisation page 8

Masquer les pouces Voir aussi pour LANGE sc200 4-20:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 26
4. Select 1 point cal.
5. Select the output mode.
Option
Description
Active
Use the current measured value during a calibration
Hold
Use the last measured value during a calibration
Transfer Transfer a specified value to the 4–20 mA input for use during a
calibration
6. Use the arrow keys to edit the displayed value so it matches the
input value from the outside device as measured by an accurate
current meter or as otherwise determined. Push the ENTER key.
The display shows the Current, Input and Cal values.
7. Push the ENTER key to continue.
If the calibration passes, the display shows a Calibration passed
message, along with the Offset and Slope values. Push the ENTER
key to continue and return to the Calibrate menu screen.
If the calibration fails, the display shows a Calibration failed
message. Refer to
Troubleshooting calibration failure
for troubleshooting suggestions.
To repeat the calibration, push the ENTER key.
To restore the previous calibration, push the BACK key. Then push
the ENTER key or wait for the controller to return to the Calibrate
menu screen.
Perform a 2 point calibration
1. From the Settings Menu, select Sensor Setup.
2. Select Calibrate.
3. Select Cal Options.
a. Select the Current option to perform a calibration using current
(mA) values, or
b. Select the Input option to perform a calibration using
measurement values, such as psi.
c. Push the ENTER key.
4. Select 2 point cal.
8 English
5. Select the output mode.
Option
Active
Hold
Transfer Transfer a specified value to the 4–20 mA input for use during a
6. Use the arrow keys to edit the point 1 value.
Note: The displayed point 1 value must match the point 1 value from the
outside device as measured by an accurate current meter or as otherwise
determined.
7. Push the ENTER key to continue.
The display shows the Current, Input and Cal values for cal point 1.
8. Push the ENTER key to continue.
9. Use the arrow keys to edit the point 2 value. The point 2 value should
be as far as possible from the point 1 value for better 2 point
calibration.
on page 9
Note: The displayed point 2 value must match the point 2 value from the
outside device as measured by an accurate current meter or as otherwise
determined.
10. Push the ENTER key to continue. The display shows the Current,
Input and Cal values for cal point 2.
11. Push the ENTER key to continue.
If the calibration passes, the display shows a Calibration passed
message, along with the Offset and Slope values. Push the ENTER
key to continue and return to the Calibrate menu screen.
If the calibration fails, the display shows a Calibration failed
message. Refer to
for troubleshooting suggestions.
To repeat the calibration, push the ENTER key.
To restore the previous calibration, push the BACK key. Then push
the ENTER key or wait for the controller to return to the Calibrate
menu screen.
Description
Use the current measured value during a calibration
Use the last measured value during a calibration
calibration
Troubleshooting calibration failure
on page 9

Publicité

Table des Matières
loading

Table des Matières