After Use; Accessories - GCE MediVital Manuel D'utilisation

Masquer les pouces Voir aussi pour MediVital:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 27
characteristics of the combination valve stated in appendix Nr. 1. To assure sufficient
performance (pressure and flow characteristics of the combination valve ) the medical
device should not be used if pressure indicator enters the red zone.
Sudden opening could result in a danger of fire or explosion arising from oxygen
pressure shocks. Insufficient opening of the shut-off valve could reduce actual flow
delivered.
If a pressure outlet as well as a flow outlet are part of the combination valve do not
use them simultaneously, especially if pressure in the cylinder is below 50 bar, in could
adversely affect the outlet parameters of the combination valve .

6.2.3 AFTER USE

Close shut-off valve (clockwise). Do not use excessive torque(max. closing torque is 5 Nm).
Vent pressure from the connected devices.
Disconnect all connected devices from user outlets.
Set flow control head on "0" (if included).

ACCESSORIES

7.
Accessories connectable to flow outlet:
hose connected with mask, cannula or humidifier.
Accessories connectable to pressure outlet:
low pressure hose (working pressure >10 bar), flowmeters, Venturi suction ejectors, lung
ventilators.
Other user accessories:
bed hanger, humidifier holder.
Accessories for filling stations:
filling adaptor.
Before connecting any accessory or medical device to the combination valve, always
check that they are fully compatible with connection features & performances of the
product.
CLEANING
8.
Remove dirt with a soft cloth damped in oil free soap water & rinsed with clean water.
Disinfection can be carried out with an alcohol-based solution (with damped wipes).
If other cleaning solutions are used, check that they are not abrasive and they are compatible
with the product materials (including labels) and gas (convenient cleaning solution - i.e.
Meliseptol)
Do not use cleaning solutions containing ammonia!
Do not expose to water or any other liquid.
Do not expose to high temperature (such as autoclave).
To apply the cleaning solution do not spray it as the spray may enter into the inner parts
of combination valve and cause contamination or damage.
Do not use pressure wash as it could damage or contaminate the combination valve .
If the inner parts of the combination valve have been contaminated do not continue
to use the combination valve
service.
under any circumstances. It must be withdrawn from
11/104
EN

Publicité

Table des Matières
loading

Table des Matières