6.2.5. Use of product through the pressure outlet.
Ensure that the flow control knob is on the ZERO position (if any).
Ensure the accessory IS NOT connected to the pressure outlet.
Slowly open the cylinder valve by turning the hand wheel in anticlock-
wise direction approx 1 to 1½ turns.
Sudden opening of the cylinder valve could result in a danger of fire or
explosion arising from oxygen pressure shocks. Insufficient opening of
the cylinder valve could reduce actual flow delivered.
Connect the accessory to the pressure outlet.
After completion of the therapy
Turn off the cylinder valve by turning the hand wheel in a clockwise direc-
tion to the stop position. Do not use excessive force.
Vent gas pressure from downstream equipment.
Disconnect the male QC probe from the pressure outlet.

6.3. AFTER USE

Turn off the cylinder valve by turning the hand wheel in a clockwise direc-
tion to the stop position. Do not use excessive force.
Reset the flow control knob on the "ZERO" position – when the gas vent-
ing has ceased (valid for version with flow-metering device only).
Ensure the pressure indicator does not show any residual pressure.
Remove connections from user outlets.
Refit pressure outlet and flow outlet protection caps. Before refitting the
caps, ensure they are clean.
7 .
CLEANING
Remove dirt with a soft cloth damped in oil free soap water & rinsed with
clean water. Disinfection can be carried out with an alcohol-based solution
(with damped wipes). If other cleaning solutions are used, check that they
are not abrasive and they are compatible with the product materials (in-
cluding labels) and gas (convenient cleaning solution - i.e. Meliseptol)
Do not use cleaning solutions containing ammonia!
Do not expose to water or any other liquid.
Do not expose to high temperature (such as autoclave).
To apply the cleaning solution do not spray it as the spray may enter into
the inner parts of product and cause contamination or damage.
Do not use pressure wash as it could damage or contaminate the pro-
duct.
If the inner parts of the product have been contaminated do not con-
tinue to use the product under any circumstances. It must be withdrawn
from service.