6.2.2. AFTER USE

Turn off at the shut-off valve by turning the hand wheel in a clockwise direction to stop
position. Do not use excessive force.
Ensure that the fl ow control knob is on ZERO position (if any).
Disconnect all downstream equipment from user outlets.
Refi t pressure outlet and fl ow outlet protection cap. Before refi tting the caps, ensure they are
clean.
7. ACCESSORY
BED HANGER
Bed hanger is used with a guard. It is designed for suspension of medical package – cylinder
and combination valve with the guard – onto the patient's bed. The maximum hanger load is
with the use of a 5-litre cylinder.
GCE FILLING ADAPTORS
GCE recommends using GCE fi lling adaptors; Non GCE fi lling adaptors can damage the non-
return valve in the fi lling port and impact the safety of the product.
8. CLEANING
Remove dirt with a soft cloth damped in oil free soap water & rinse with clean water. Used
cleaning agents shall comply with oxygen cleanliness standards.
Disinfection can be carried out with an alcohol-based solution (spray or wipes).
If other cleaning solutions are used, check that they are not abrasive and they are compatible
with the product materials (including labels) and gas (convenient cleaning solution - i.e.
Meliseptol).
Do not use cleaning solutions containing ammonia!
Do not expose to water or any other liquid.
Do not expose to high temperature (such as autoclave).
To apply the cleaning solution do not spray it as the spray may enter into the inner parts
of combination valve and cause contamination or damage.
Do not use pressure wash as it could damage or contaminate the combination valve.
If the inner parts of the combination valve have been contaminated do not continue to
use the combination valve under any circumstances. It must be withdrawn from service.
9. SERVICE, PRODUCT LIFE TIME AND MAINTENANCE
9.1. SERVICE PERIOD AND PRODUCT LIFE TIME
SERIAL NUMBER AND MANUFACTURE DATE
A nine-fi gure serial number stamped on valve body consists of the data as follows:
YY MM XXXXX
YY: year of manufacture
MM: month of manufacture
XXXXX: sequence number of product
For example: Serial number 050300521 indicates the valve manufactured in 2005, in March,
with sequence number 521.
An eleven-fi gure serial number stamped on valve body consists of the data as follows:
YYYY MM XXXXX
YYYY: year of manufacture
MM: month of manufacture
XXXXX: sequence number of product
For example: Serial number 20050300521 indicates the valve manufactured in 2005, in March,
with sequence number 521.