Recommended separation distances between portable and mobile RF
communications equipment and the Smoke Check
The Smoke Check is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The customer or the user of the Smoke Check
can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and the
Smoke Check as recommended below, according to the maximum output power of the
communications equipment
Rated Maximum
Output Power of
Transmitter in
Watts (W)
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the
recommended separation distance d in meters (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the maximum
output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency
range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and reflection from structures objects and people
.
Separation Distance in Meters (m) according to Frequency of
Transmitter
150 KHz to 80 MHz
d = 1.2√P
0.12
0.38
1.2
3.8
12
80 MHz to 800 MHz
d = 1.2√P
0.12
0.38
1.2
3.8
12
62
800 MHz to 2.5 GHz
d = 2.3√P
0.23
0.73
2.3
7.3
23.3