Electromagnetic Compatibility (EMC) to
EN60601-1-2:2007.
WARNING: use of portable phones or other radio frequency (RF)
emitting equipment near the system may cause unexpected or adverse
operation.
The Smoke Check monitor has been tested to EN 60601-1-2-2007, regarding
its ability to operate in an environment containing other equipment
electrical/electronic equipment (including other medical devices)
The purpose of this testing is to ensure that the Smoke Check monitor is not
likely to adversely affect the normal operation of other such equipment and
that other such equipment is not like to adversely affect the normal operation
of the Smoke Check monitor.
Despite the testing of the Smoke Check that has been undertaken, normal
operation of the Smoke Check can be affected by other electrical/electronic
equipment and portable and mobile RF communication equipment.
As the Smoke Check is medical equipment, special precautions are needed
regarding EMC (electromagnetic compatibility).
It is important that the Smoke Check monitor is configured and installed/put
into service in accordance with the instructions/guidance provided herein and is
used only in the configuration as supplied.
Changes or modifications to the Smoke Check may result in increased
emissions or decreased immunity of the Smoke Check in relation to EMC
performance
The Smoke Check should not be used adjacent to or stacked with other
equipment. If adjacent or stacked use with other equipment is necessary,
the Smoke Check and the other equipment should be observed / monitored,
to verify normal operation in the configuration in which it will be used.
58