TV 141 NAVIGATOR VENT VALVE KIT INSTALLATION PROCEDURE
OVERVIEW
The "TV 141 Navigator Vent Valve" kit, consisting of a
control unit and valve, is a complete unit for automatic
pump venting when the pump is switched off or during
a power failure. The valve is normally open during
power off conditions, is electromechanically activated
and secured (Viton-sealed) by means of an M8 thread
with related O-ring on the high vacuum port. The air
input into the valve is filtered by means of a specific
filter located on the valve's own air inlet.
The control unit is powered by the Turbo-V controller
which is not suitable for rack mounting.
The control unit is activated with a preset delay of
approximately 0.8 seconds to avoid undesired venting
during a temporary power failure and to allow the
closure of the system valves before venting.

TECHNICAL CHARACTERISTICS

Control Unit
Input voltage
- power (max)
Output voltage (max)
- power (max)
Delay
Operating temperature
Storage temperature
Connection Cables
Input
Output from valve
24 Vdc ± 10%
2.5 W
24 Vcc ± 10%
1.2 W
Approx. 0.8 seconds
0 to 40 °C
-20 to 50 °C
120 mm (4.72 inches)
200 mm (7.87 inches)
Vent valve
Valve status
Vacuum connection
Air intake filter
Hole dimension
Pressure ranges
Air intake adapter
Leak rate
Life span
Input voltage
Power
Bakeout temperature
Weight
The
following
figure
dimensions.
Dimensions in mm
7

INSTALLATION PROCEDURE

Normally open
(closed when power is
applied)
M8 thread
Sintered bronze
0.8 mm (0.03 inch)
-6
10
mbar to 1 bar
-7
(10
Torr to 760 Torr)
Ø 6.35 mm (¼ inch)
≤ 1x10
-7
mbar l/s
1 milion cycles
24 Vdc ± 10%
2.5 W
60 °C
140 g (0.3 lbs)
shows
the
valve's
87-900-911-01 (C)
overall