Table 14.
Guidance and manufacturer's declaration – electromagnetic immunity for
the Aerogen USB Controller system that is not life supporting
This Aerogen USB Controller system is intended for use in the electromagnetic
environment specified below. The customer or the user of the Aerogen USB Controller
system should assure that it is used in such an environment.
IEC 60601 Test
Immunity Test
Level
Electrostatic
±8 kV contact
discharge (ESD)
IEC 61000-4-2
±15 kV air
EN 61000-4-2:
2009
Electrical fast
±2 kV for power
Transient/burst
supply lines
IEC 61000-4-4
±1 kV for input/
EN 61000-4-4:
output lines
2012
Surge
±1 kV line(s) to
line(s)
IEC 61000-4-5
EN 61000-4-5:
±2 kV line(s) to
2006
earth
Voltage
<5 % Ut
dips, short
(>95 % dip in
interruptions
Ut) for 0.5 cycle
and voltage
@ 0°, 45°, 90°,
variations on
135°, 180°, 225°,
power supply
270°, 315°
input lines
IEC 61000-4-11
70 % Ut
EN 61000-4-11:
(30 % dip in Ut)
2004
for 25 cycles
<5 % Ut
(>95 % dip in Ut)
for 5 sec
Compliance
Electromagnetic
Level
Environment - Guidance
±2, 4, 6 & 8 kV
Floors should be wood,
contact
concrete or ceramic tile.
If floors are covered with
±2, 4, 6, 8 & 15
synthetic material, the
kV air
relative humidity should be at
least 30%.
±2 kV for power
Mains power quality
supply lines
should be that of a typical
commercial or hospital
±1 kV for input/
environment.
output lines
±1 kV line(s) to
Mains power quality
line(s)
should be that of a typical
commercial or hospital
±2 kV line(s) to
environment.
earth
<5 % Ut
Mains power quality
(>95 % dip in
should be that of a typical
Ut) for 0.5 cycle
commercial or hospital
@ 0°, 45°, 90°,
environment. If the user of
135°, 180°, 225°,
the Aerogen USB Controller
270°, 315°
system requires continued
operation during power mains
operation, it is recommended
70 % Ut
that the Aerogen USB
(30 % dip in Ut)
Controller system be powered
for 25 cycles
from an uninterruptible power
supply.
<5 % Ut
(>95 % dip in Ut)
for 5 sec
Aerogen
®
207