Hach FP 360 sc Manuel D'utilisation De Base page 8

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 19
Function test
C A U T I O N
Before power is applied, refer to the controller operation
instructions.
After the components are removed from the package, do a function test.
Connect the connector cable to the sensor (8-pin polarized
1.
connector) and an appropriate sc controller (5-pin polarized
connector) (refer to
Connect sensor cable on page
Apply power to the sc controller. The display is activated and the
2.
sensor goes to measurement mode.
The sensor ticks quietly and regularly.
Cover the sensor measurement window with a sheet of white paper
3.
(do not use recycled paper).
Vary the distance between the measurement window and the
4.
paper. The measurement value on the display will change
accordingly.
Note: In air, the measurement value displayed is not exactly zero due to
reflections on the window surface (refer to
page
14).
Installation
D A N G E R
Personal injury hazard. Only qualified personnel should conduct
the tasks described in this section of the manual.
N O T I C E
If the sensor is not fully inserted, sun protection is recommended
in high ambient temperatures and intense solar radiation to
protect against thermal and UV effects.
English 8
8).
Check the zero point on
Connect sensor cable
C A U T I O N
Always put the cables and hoses in a position that does not bend
or cause a trip hazard.
C A U T I O N
Before power is applied, refer to the controller operation
instructions.
Connect the polarized socket on the connector cable to the sensor
1.
plug (8-pin plug) (refer to
Push the safety sleeve on the plug (step 2).
2.
Hand-tighten the safety sleeve in position (step 3).
3.
Figure 3 Connect the sensor cable to the sensor
Figure 3
step 1).

Publicité

Table des Matières
loading

Table des Matières