Dispense/Copy; Keypad Lockout Enable/Disable; Remote Control; Remote Control Setup - Masterflex L/S 07575-00 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

DISPENSE/COPY

A first press of the DISP key results in the last entered dispense volume being displayed. The VOL annunciator will
illuminate and flash. The INC/DEC (
The START/STOP key then initiates delivery of the set volume. The amount remaining to be dispensed will be
displayed during countdown. Resetting cumulative dispensed volume can be done by pressing and holding the
TOTAL VOLUME key until zero is displayed (3 seconds). Display flashes "99999" when cumulative volume is greater
than 99999. The dispense function is exited by the FLOW key.
A second press of the DISP key causes the COPY annunciator to illuminate and flash. The START/STOP key is then
used to deliver the desired volume without the need to know the volume in specific units. A third press of the DISP key
enters the volume dispensed. The COPY annunciator stops flashing. The START/STOP key is then used to initiate
delivery of the copied volume. The number of copies dispensed is displayed after each dispense. The maximum
number of copies is 99999. The START/STOP key is used to pause the copy dispense during dispensing; copy
dispense can then be continued using the START/STOP key.
A fourth press of the DISP key results in the last entered dispense interval being displayed. The SEConds annunciator
will illuminate and flash. The INC/DEC ( ,
from 0 to 99999 seconds. The START/STOP key then initiates delivery for the set volume, with the drive automatically
initiating a new dispense after each time out. The remaining time will be displayed during countdown. The
START/STOP key is used to stop the dispense cycle. A time of 0 seconds (default) will require initiation of each dispense
through the START/STOP key or the Remote START/STOP contact closure. Pressing the DISP key a fifth time exits this mode.

KEYPAD LOCKOUT ENABLE/DISABLE

Press/hold FLOW. After five (5) seconds, display changes to all dashes. While holding FLOW, press PRIME five (5) times.

REMOTE CONTROL

Selectable input (0–20 mA, 4–20 mA, 0–10V DC )
±0.5% linearity control
±50V common mode range with earth ground
START/STOP; CW/CCW; PRIME via contact closure

REMOTE CONTROL SETUP

1. Place the power switch in the off position.
CAUTION:
Power must be turned off before connecting the external remote control cable to prevent
damage to the drive.
2. Connect the cable from the external remote control to the mating receptacle on the rear panel.
3. Select type of remote control input and output required as follows:
a. Press and hold the MODE control while turning the power switch to the on position. After two seconds, release
the MODE control. The initial display will show: "inP". After two seconds, the display shows either 0–20 or 4–20.
NOTE:
Press the INC ( ) or DEC ( ) arrow keys to select between 4–20 and 0–20 for current loop control.
b. Press the MODE control again. The initial display will show: "out". After two seconds, the display shows either
0–10, 0–20, or 4–20.
NOTE:
Press the INC ( ) or DEC ( ) arrow keys to select between 4–20, 0–20, or 0–10 for current loop
or voltage output.
4. Press the MODE control to select mode of operation. The LEDs indicate the selected mode. Select either mA or V.
6
,
) keys are used to change the dispense volume, if desired.
) keys are used to change the time interval between dispenses, if desired,