Guidance and manufacture's declaration-electromagnetic emission
The Fingertip Oximeter is intended for use in the electromagnetic environments specified
below, the customer or the user of the Fingertip Oximeter should assure that it is used in such
an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
Note: U
is the a.c. mains voltage prior to application of the test level.
T
IEC 60601
test level
± 6 kV contact
± 8 kV air
± 2 kV for power
supply lines
± 1 kV for input/
output lines
± 1 kV line(s)
to line(s)
± 2 kV line(s)
to earth
<5% U
T
(>95% dip in U
)
T
for 0.5 cycle
<40% U
T
(60% dip in U
)
T
for 5 cycle
<70% UU
T
(30% dip in U
)
T
for 25 cycle
<5% U
T
(>95% dip in U
)
T
for 5 sec
3A/m
Compliance
Electromagnetic
level
environment-guidance
± 6 kV contact
Floors should be wood, concrete
± 8 kV air
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be
at least 30%.
N/A
N/A
N/A
N/A
N/A
N/A
3A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.