Guidance and manufacturer's declaration - electromagnetic immunity for
the Aerogen Solo nebuliser system that is not life supporting
This Aerogen Solo nebuliser system is intended for use in the electromagnetic
environment specified below. The customer or the user of the Aerogen Solo nebuliser
system should assure that it is used in such an environment.
IEC 60601 Test
Immunity Test
Level
Conducted RF
3 Vrms outside
IEC 61000-4-6
industrial,
EN 61000-4-6:
scientific and
2014
medical (ISM) and
amateur radio
bands. 6 Vrms in
ISM and amateur
radio bands
150 kHz to 80 MHz
Note:
Compliance
Electromagnetic
Level
Environment - Guidance
Portable and mobile RF communications
equipment should be used no closer to
any part of the Aerogen Solo nebuliser
system, including cables, than the
recommended separation distance
calculated from the equation applicable
to the frequency of the transmitter.
Recommended Separation Distance
10 Vrms
d = [1.17] √P