Table 10.
Guidance and manufacturer's declaration – electromagnetic immunity for
the Aerogen Solo nebuliser system that is not life supporting
This Aerogen Solo nebuliser system is intended for use in the electromagnetic
environment specified below. The customer or the user of the Aerogen Solo nebuliser
system should assure that it is used in such an environment.
IEC 60601 Test
Immunity Test
Level
Electrostatic
±8 kV contact
discharge (ESD)
IEC 61000-4-2
±15 kV air
EN 61000-4-2:
2009
Electrical fast
±2 kV for power
Transient/burst
supply lines
IEC 61000-4-4
±1 kV for input/
EN 61000-4-4:
output lines
2012
Surge
±1 kV line(s) to
line(s)
IEC 61000-4-5
EN 61000-4-5:
±2 kV line(s) to
2006
earth
Voltage
<5 % Ut
dips, short
(>95 % dip in
interruptions
Ut) for 0.5 cycle
and voltage
@ 0°, 45°, 90°,
135°, 180°, 225°,
variations on
power supply
270°, 315°
input lines
IEC 61000-4-11
70 % Ut
EN 61000-4-11:
(30 % dip in Ut)
2004
for 25 cycles
<5 % Ut
(>95 % dip in Ut)
for 5 sec
Compliance
Level
±2, 4, 6 & 8 kV
contact
±2, 4, 6, 8 & 15
kV air
±2 kV for power
supply lines
±1 kV for input/
output lines
±1 kV line(s) to
line(s)
±2 kV line(s) to
earth
<5 % Ut
(>95 % dip in
Ut) for 0.5 cycle
@ 0°, 45°, 90°,
135°, 180°, 225°,
270°, 315°
70 % Ut
(30 % dip in Ut)
for 25 cycles
<5 % Ut
(>95 % dip in Ut)
for 5 sec
Aerogen
®
Electromagnetic
Environment - Guidance
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should be at
least 30%.
Mains power quality
should be that of a typical
commercial or hospital
environment.
Mains power quality
should be that of a typical
commercial or hospital
environment.
Mains power quality
should be that of a typical
commercial or hospital
environment. If the user of
the Aerogen Solo nebuliser
system requires continued
operation during power mains
operation, it is recommended
that the Aerogen Solo
nebuliser system be powered
from an uninterruptible power
supply or battery.