2.1 Connecting in four-wire and six-wire configurations

When load cells using a six-wire configuration are connected to amplifiers with
a four-wire configuration, the sense leads of the load cells must be connected
to the corresponding supply leads: marking (+) with (+) and marking (−) with
(−).
6-wire cable connection
Schematic diagram of a Pancon plug (CE100F26-6), 6-pin
2.2 Cable protection
Cable protection system*, comprising:
Unscrew the M10 sleeve and replace
with an M10 threaded tube coupling
Even without a cable protection system, standard load cells can achieve
IP68/IP69K degree of protection. In applications where the PUR cable of the
load cell can be chemically or mechanically attacked or destroyed, extra cable
protection can be provided by means of standard cable protection systems.
A3010−1.0 en/de/fr
Shield (yellow) = cable shield
connected to load cell body
Contact 1 (white) = Signal (+)
Contact 2 (red) = Signal (−)
Contact 3 (black) = Excitation (−)
Contact 4 (blue) = Excitation (+)
Contact 5 (green) = Sense (+)
Contact 6 (gray) = Sense (−)
blue marking
(provided by customer)
Screwed cable gland
Flexible cable protection tubing*
* Supplier such as Flexicon,
e-mail: info@icp-solutions.de
9