Debouncing is dependent on the function set (IMD command), but not on the
input used (cable 1 or cable 2).
The differences between the various inputs must be noted for the electrical
data:
Reference potential
Low level
High level
Input resistance

4.2 Data of outputs

The output driver of the FIT
driver (static switch). The circuit is potentially isolated from the FIT
measurement electronics and must be supplied via the external operating
voltage UB 2. When an output is activated, a positive level is given on the
corresponding output line, voltage is dependent on UB 2. The consumer must
be switched between the output and the minus pole of the power supply.
Voltage
UB 1
supply
measuring
GND 1
circuit
Example for output configuration, the outputs are not set to specific
consumers (lamps, relays).
A1779-2.0 en/de/fr/es
Trigger input (cable 1)
GND 1
0...1 V
3...12 V
10 kΩ
R
switch outputs is a PLC compatible HighSide
FIT
UB 2
Out 1
Out 2
Out 3
Out 4
GND 2
Inputs IN1, IN2
GND 2
0...6 V
10...30 V
> 3 kΩ
R
+
Voltage
supply
for ext.
consumer
17