How To Change Parameter Values - Kannad PROVBIO-B Manuel Utilisateur

Flottant
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

USER MANUAL / MANUEL UTILISATEUR
• 44.4444 is the transmittance,
• 5555 6666 7777 are backscattering, CDOM and chlorophyll.
5.8.5.5. Test Hydraulic Pump
To activate the pump for one second, send the command:
!P 10
Listen for the pump running.
5.8.5.6. Test Hydraulic Valve
To activate the valve for one second, send the command:
!E 10
Listen for the actuation of the valve.
5.8.5.7. Test Iridium Subsystem
To test the Iridium transmitter, send the command:
!SE
This command executes a Loc GPS / transmission / reception phase and will cause PROVBIO-B to
transmit several messages. They are technical messages, the format of which is described in
IRIDIUM DATA PACKETS FORMATS page
Put the magnet back in place to stop the transmission.
You have now completed the functional tests. Ensure the magnet is in place.

5.9. How to Change Parameter Values

Your PROVBIO-B float has been pre-configured at the factory. Should you wish to change any of the float's
parameters, you must follow the instructions provided below.
Connect the PC to the float using the RS232 interface cable
You can read parameter values at any time by sending the PM command as explained in
to Read Parameter Values page
The commands for changing the values of the mission parameters are of the form:
!PM X Y
where X identifies the parameter and Y provides its new value.
Remember that you may use the parameter number in place of X (see
parameter numbers).
For example, to change the number of cycles to 150 send the command:
!PM 01 150
PROVBIO-B will respond:
<PM1 150>
NOTE:
PROVBIO-B will always respond by confirming the present value of the parameter. This
is true even if your attempt to change the parameter's value has been unsuccessful, so
you should observe carefully how PROVBIO-B responds to your commands.
You can set the time on the float's internal clock by sending the command:
!TI YYYY/MM/DD h:m:s
For example, if you send the command:
!TI 2008 07 13 15 20 00
PROVBIO-B will respond:
<TI a2008 m7 j13 h15 m20 s0>
PROVBIO-B
20. Use your e-mail to receive the message.
13.
(section 5.3. Connecting the PC page
Table 1: page 9
PAGE : 17 / 72
DATE : 26/03/2008
INDEX : A
REF. : DOC08018
section 6.
12).
section 5.5. How
for a summary of all

Publicité

Table des Matières
loading

Produits Connexes pour Kannad PROVBIO-B

Table des Matières