6. Error messages

During calibration
On display
Asymmetry Potential
Sensitivity
Sensitivity
Buffer solutions at
different temperature
Unstable reading
more than 1 min
SAME BUFFER
SOLUTIONS
UNKNOWN
BUFFER SOLUTION

Other messages

On display
OUT OF RANGE
°C OUT OF RANGE
si627 P_HACH_LANGE_Ed.0510.indd ANG:19
si627 P_HACH_LANGE_Ed.0510.indd ANG:19
Problem
Electrode asymmetry
potential > 60 mV.
> 60 mV
Electrode slope < 85 % of
theoretical value.
< 85 %
Electrode slope > 103 % of
theoretical value.
> 103 %
Buffers temperature.
Unstable reading for more
than 60 seconds.
1st and 2nd buffer solutions
are the same.
Incorrect buffer solution.
Problem
Measured value out of
range.
Measured temperature out
of range.
Possible causes
Electrode "worn out".
Buffer solutions contaminated.
Electrode "worn out".
Buffer solutions contaminated.
Electrode "worn out".
Buffer solutions contaminated.
The temperature difference between both buffer
solutions used in calibration is > 3 °C.
Press ESC to quit.
Electrode defective or dirty, see miscellaneous solu-
tions. Variation of buffer solution temperature.
(i.e. when stored previously in the refrigerator).
Same standard was used for both calibration steps.
Electrode is broken or cable defective.
Press ESC to quit.
Same standard was used for both calibration steps.
Electrode is broken or cable defective.
Press ESC to quit.
Possible causes
pH value < -2 or > 16, mV < –1500 or > 1500.
The current output will be frozen at 22 mA.
(Alarm).
Pt 1000 broken or temperature < –20 or >
.
150°C
The current output will be frozen at 22
mA. (Alarm)
19
20/5/10 10:50:49
20/5/10 10:50:49