4.2.4 XSE balance settings for Antistatic Kit
The arrow buttons can be used to page forward or back to a menu page.
1 Switch on the balance [ ].
2 Press [ ].
3 Choose ErgoSens and tap the button [Define].
ð A selection window appears.
4 Choose ErgoSens 1 (Aux1) or ErgoSens 2 (Aux2) and tap on the associated button.
ð A selection window appears.
5 Tap in the scroll bar to select [AntiStatic Kit].
6 To save the settings, tap [OK].
7 Choose Function Keys and tap the button [Define].
ð A selection window appears.
8 Tap [Ionizer].
9 To save the settings, tap [OK].
ð The function key [Ionizer] is displayed in the application at the bottom of the display.
ð Indicator light turns on and Status-Icon [ ] appears in the display when the Ionizer is working.
10
Installation and Putting into Operation​ ​
Compact Antistatic Kit