7. Starting, operation
Mechanical connection
For pressure switches with straight thread the sealing ring is included in delivery.
When mounting the instrument, ensure that the sealing faces of the instrument and
the measuring point are clean and undamaged.
Screw in or unscrew the instrument only via the flats using a suitable tool and the
prescribed torque. The appropriate torque depends on the dimension of the pressure
connection and on the sealing element used (form/material). Do not use the case as
working surface for screwing in or unscrewing the instrument.
When screwing the pressure switch in, ensure that the threads are not jammed.
Types of sealings
according to DIN 3852-E
NPT is a self-sealing thread type.
Betriebsanleitung / Operating manual / Instructions de service / Instrucciones de uso
For the assembly the
following moments may
not be exceeded:
G 1/8 = 15 bis 20 Nm
G 1/4 = 20 bis 25 Nm
G 1/2 = 40 bis 50 Nm
according to NPT