TABLE 2
Immunity test
Electrostatic discharge
(ESD)
IEC 61000-4-2
Electrical fast transient/
burst
IEC 61000-4-4
Surge IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations on
power supply input lines
IEC 61000-4-11
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
NOTE: UT is the a.c. mains voltage prior to application of the test level.
18
DECLARATION - ELECTROMAGNETIC IMMUNITY
IEC 60601 test level
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
± 2 kV for power supply lines
± 1 kV for input/output lines
± 0.5kV, ± 1 kV line(s) to lines
± 0.5kV, ± 1 kV, ± 2 kV line(s) to
earth
0 % UT; 0.5 cycle At 0°, 45°, 90°,
135°, 180°, 225°, 270°and 315°
0 % UT; 1 cycle and
70 % UT; 25/30 cycles
Single phase: at 0°
0 % UT; 250/300 cycles
30 A/m
Compliance level
±8 kV contact
±2 kV, ±4 kV, ±8
kV, ±15 kV air
Not applicable
Not applicable
Not applicable
30 A/m