* EVT [WARM_RESET_SYSTEM_RESTART]
to inform that a warm reset system restart is happened and the HV
output, that was in the ON state before the discharge, is
automatically restarted.
After these five lines, the normal report sequence starts, accordingly
to the report type chosen.
Enable – Disable Remark Lines
These five lines are sent also if the EEPROM's parameter n. 30 are set
to "0" = Disable Remark Lines.
If the EEPROM's parameter n. 30 is set to "1", all the remark lines,
the lines that begin with the symbol "*", are sent.
For EEPROM's parameter n. 30 = "0", all the remark lines are NOT
sent, except the first six lines at the main power on reset (cold reset)
and the first five lines after an ion pump discharge reset (warm
reset) sent anyway.

Continuous and NON continuous Report Mode

IPCU is able to send report only on request. To do this, it is necessary
to set the report mode with the time rate = 0.
After the "enter", the continuous report mode stops immediately.
According to the report type and mode, the unit sends one
instantaneous report each time a "Request Report" (RR) command is
received.
To return in the continuous report mode it is necessary to send a
new "Report Setting" (RT) command with time rate ≠ 0.
After the "enter" the continuous report mode starts immediately.
Ion Pumps Control Unit User Manual / 87-900-124-01
Technical Information
(C)
5
PIN Functions
91/104