Télécharger Imprimer la page

Analog Recorder Outputs; Serial Operation - Agilent Technologies 629-5001 Notice De Mode D'emploi

Publicité

Analog Recorder Outputs

Figure 9
Analog Recorder Output:
Ch1: Pin 11 (+), Pin 24 (0V)
At Pin 25 a reference voltage corresponding to the 2 recorder outputs full-scale
NOTE
value (5 V +/- 5 %) is given. This reference allows getting a more precise
reading from the 2 recorder outputs. It allows the scaling of the actual reading
to the real value of the full scale.

SERIAL Operation

The Controller works on RS232 DCE Mode (like a Modem with DB9
FEMALE connector).
The Host/Terminal/Personal Computer works always on RS232 DTE
Mode (like a PC with DB9 MALE connector).
The Cable Interconnections are 1 to 1 signals (DB9 Male/Female
Standard CABLE).
Ion Pumps Control Unit User Manual / 87-900-124-01
Technical Information
ION Pump Controller Description
Ch2: Pin 12 (+), Pin 24 (0V)
(C)
5
81/104

Publicité