Serial Operation - Agilent Technologies 629-5001 Notice De Mode D'emploi

Publicité

4
Installation Procedure
Operating Modes
If the connection between the IPCU and the pump is done by means of an HV
NOTE
Cable with Safety Interlock, the cable must be connected to the pump F/T and
the small banana connector on the cable controller side must be inserted in the
corresponding "HV Cable Interlock" plug on the unit rear panel in order to be
able to switch on the HV.
If the connection is done with a single pole HV Cable without the
Safety Interlock, ensure that the jumper provided on the unit rear
panel is inserted in the "HV Cable Interlock" plug.

SERIAL Operation

The SERIAL is set in order to operate by connecting the controller to an external
NOTE
PC that uses the "HYPER TERMINAL" or equivalent program (for example a
proprietary user program) running under Windows Operating System or other.
To operate in serial mode, connect the serial port to the controller "Serial"
connector on the rear panel, run the Windows "HYPER TERMINAL" program on
the PC with the following default settings:
This way of operation will allow to verify the functionality of the unit in the
"SERIAL" operating mode, give the commands and get all the operating data.
In the serial operation, depending on the user environment, the O.S. used must
NOTE
have real time capability.
See the "Technical Information" paragraph for details.
68/104
COM1
9600 baud
No parity
8 data bits
1 stop bit
Ion Pumps Control Unit User Manual / 87-900-124-01