Technical Information
Pin Functions
These six lines are sent also if the EEPROM's parameter n. 30 are set
to "0" = Disable Remark Lines.
After these six lines, the normal report sequence starts, accordingly
to the report type chosen notifying the user about the values of
voltage, current and power, faults and events status for each channel.
When an ion pump discharge happens, the unit performs a warm
system start reset automatically bringing to 0 V the HV to stop the
discharge, then it sends again the first five remarking lines with the
fifth changed to:
* EVT [WARM_RESET_SYSTEM_RESTART]
to inform that a warm reset system restart is happened and the HV
output, that was in the ON state before the discharge, is
automatically restarted.
After these five lines, the normal report sequence starts, accordingly
to the report type chosen.
Enable – Disable Remark Lines
These five lines are sent also if the EEPROM's parameter n. 30 are set
to "0" = Disable Remark Lines.
If the EEPROM's parameter n. 30 is set to "1", all the remark lines,
the lines that begin with the symbol "*", are sent.
For EEPROM's parameter n. 30 = "0", all the remark lines are NOT
sent, except the first six lines at the main power on reset (cold reset)
and the first five lines after an ion pump discharge reset (warm
reset) sent anyway.

Continuous and NON Continuous Report Mode

IPCU is able to send report only on request.
To do this, it is necessary to set the report mode with the
time rate = 0.
90/106
Ion Pumps Control Unit (No Battery) User Manual / 87-900-123-01 (B)