g)
ALTERNATING LOW PRESSURE MODE INDICATOR Adjacent cells within mattress will be
alternately inflated and deflated in this mode. The indicator adjacent to the symbol will illuminate
while Alternating Low Pressure mode is active. Note: Alternating Low Pressure mode is
temporarily suspended in Max Inflate mode.
h)
SOFT (−) BUTTON decreases the inflation level of the bed while Control Unit is in Manual mode
only. Pressing once will decrease the setting by one comfort level.
i)
COMFORT CONTROL INDICATOR displays the current inflation level set point.
Note: An audible alarm will activate and the comfort level indicator(s) will flash if a set point not
achieved alarm is activated.
j)
FIRM (+) BUTTON increases inflation level of the bed while in Manual mode only. Pressing once
will increase the setting by one comfort level setting (~2.5mmHg).
k)
MAX INFLATE BUTTON AND INDICATOR selects/deselects the Max Inflate mode (maximum
mattress firmness). The indicator adjacent to the button will illuminate while Max Inflate mode is
active. After 30 minutes, Max Inflate mode will cancel.
l)
LOCKOUT BUTTON AND INDICATOR activates or cancels Lockout mode. Press and hold for
three seconds to activate or cancel mode. The indicator adjacent to the button will illuminate while
Lockout mode is active. In Lockout mode, only the Max Inflate button is operational. All other
buttons are temporarily inactive.
m)
ALARM SILENCE BUTTON when this button is momentarily pressed it silences indefinitely any
existing audible alarm.
n)
ALARM INDICATOR flashes indefinitely when an alarm exists. If no alarm conditions are present
this indicator will extinguish.
o)
POWER FAIL INDICATOR flashes synchronously with an audible alarm when the power supplied
from line cord is interrupted while the Control Unit is operating. To avoid an inadvertent power fail
alarm, place the unit in Standby mode before disconnecting power from Control Unit.