A.4 Installation Of Camera Cables; A.5 Installation Of Video Camera With C-Mount Interface; A.6 Handling The Slit Lamp When Fitted With Video Equipment - Haag-Streit BD 900 Mode D'emploi

Table des Matières

Publicité

A.4
Installation of camera cables
Secure the cable to the camera (100), lead it through the table
(102) and connect it to the camera control unit. On the slit lamp
the cable is fixed on the microscope arm (99) and additionally
on the rail cover with the cable guide (101).
A.5
Installation of video camera
with C-mount interface
Unscrew the cover (103) on the C-mount interface. First screw
the adapter (104) on the camera and then affix it with 3 set
screws (105).
Image position on the monitor
Note the position of the camera! The camera must be oriented
parallel to the microscope arm and must be mounted in the way,
that the picture is correctly displayed on the screen.
In the case of misaligned images on the screen, the camera
should be rotated until the correct position image is dis-
played.
Image focus on the monitor
Needs not to be adjusted, the interface has been designed so
that optimal image focus is set automatically!
Breath shield
Depending on camera model, a distance screw must be built
in (106), thus the breath shield can be fixed correctly.
A.6
Handling the slit lamp when fitted with
video equipment
To achieve the same focus through both eye-pieces and
monitor, the eye-pieces must be exactly adjusted with the
focus test rod.
When coupling the accommodation facility with convergent
viewing, the normal refraction often has to be corrected by -1
to -2 dpt on the eye-piece.
99
100
101
102
103
104
105
106
1500.7200631.04040
© Haag-Streit AG, CH-3098 Koeniz, Switzerland