2.3 Electrical connection

- Before connecting the unit, note the specifications on the name plate.
- The unit may be connected only by a certified electrician in compliance with country-specific and
local regulations.
- The unit is delivered with a connection cable of approx. 1.5 m without plug. If the terminal line has
to be replaced for structural reasons, a type H07BQ-F cable is to be used.
- An all-pole electric separator with a contact distance of at least 3 mm is to be connected in series
near the unit by the customer. The connection cable shall only be replaced or extended by the
manufacturer or an authorised Salvis Service.
- Each unit requires its own power connection and fuse.
- The unit must be included in the potential equalization system.
The M6 thread for the connection terminal is located behind the guide column
.
2.4 Dimensioned drawing of the table model
512
204
162
450
S A L300
227
450
500
Installation note:
Never carry the unit by the reflector!
Lay the cable upward neatly in a cable conduit.
90
15
from above
4 M6 threads
name plate
40
273
10
7 / 9
.
130
80
600
SA L301
490
550
600