4. The Return icon takes you out of Setup. The last shown settings will become the new
defaults.
Заказывайте у нас на сайте ww.euromed.in.ua
The first system setup screen (Model 5010 and 5015)
Note: Model 5000 does not have Keratometry, Rounding (K)
and PK readings options
Rounding
Choose 0.12 or 0.25 for rounding refraction results. The default is 0.25 diopters.
Vertex
Choose a vertex setting of 0, 10.5, 12.0, 13.5, 15.0 or 16.5 mm. The default is 13.5.
Language
The available languages are English, Danish, Dutch, French, German, Italian, Norwegian,
Portuguese, Spanish, and Swedish.
Eye Monitor
The position of the eye monitor on the refraction screen can alternate from right to left,
depending on the eye being refracted, Alternate, or remain fixed on the left side if the
operator prefers, Stationary.
A third position of the eye monitor, Full Screen, allows a larger image of the eye to be
displayed upon target detection for refraction. The default is Full Screen.
Brightness
Adjust screen brightness from minimum to maximum, 1 to 5. The default is 5.
Keratometry
Choose mm or dK. The default is dK (Model 5010 and 5015).
Rounding (K)
Choose 0.12 or 0.25 for rounding keratometry results.
The default is 0.25 diopters. (Model 5010 and 5015)
Peripheral
Choose to measure the right (Rt) and left (Lt) peripheral
K Readings
points; or the right, left, superior, and inferior points (All).
The default is All. (Model 5010 and 5015)
Zeiss Humphrey Systems Acuitus User Manual
PN 50648 Rev A June 2001
4