9.2.3
Déroulement d'une mesure
L'illustration suivante représente le déroulement d'une opération de mesure.
MCP Operator
5: return status()
20: return status()
MCP Controller
1: dispense desired volume
2: press 'START'
3: measurement started (A10 8_B<CRLF>)
4: make COM port reservation()
7: return channel #1 weight (e.g. 0.00015 <TAB>)( )
9: return channel #2 weight (e.g. 0.00014<TAB>)( )
10: ...channels #3..#7...
12: return channel #8 weight (e.g. 0.00013 <TAB>)( )
14: return channel #1 check weight (e.g. 0.0 0012<TAB>)()
15: return weight unit ( g<TAB>)()
16: return place holder ( _<TAB>)()
17: return place holder ( _<TAB>)()
18: return serial number ( 1113153280<CRLF>)( )
19: measurement done (A10 8_A<CRLF>)( )
21: display status 'READY'()
Annexe: MCP et ordinateur hôte
MCPxxx
6: measure channel #1()
8: measure channel #2()
11: measure channel #8()
13: measure channel #1 again()
37