NOTES

-
The display shows the "Err-Lo" message when one enters a lower limit greater than the upper limit.
-
The weight check is enabled with a weight greater than 0.
-
The high limit and low limit set in the weighing model are saved in the scale.

6. SET-UP

6.1 NUMBER OF DIVISIONS AND ZERO TRACKING

Press ENTER and CALI. keys at length:
-
The display shows the set number of divisions "d XXXX": press
NWLH/NWSH models it is possible to select also 15000 or 30000 maximum divisions.
-
By pressing the ENTER key the display shows the set zero tracking value "ZEr X.X", in other words, scale's
thermal drift compensation parameter; the set value corresponds to the number of divisions which is cleared.
Press
to select 0.0 (disabled), 0.5, 1.0, 1.5, 2.5 or 3.5.
-
Press ENTER to confirm.

6.2 GRAVITY OF CALIBRATION AND USE ZONES

a) Keep pressed the ENTER key for about 4 seconds: the display will show "CAL_G".
b) Press
: input the gravity value of the calibration zone, by increasing the digit with the
to set the successive digit by pressing ENTER.
c) Once the last digit is confirmed the display will show "CAL_G".
d) Press
: input the gravity value of the use zone, by increasing the digit with the
the successive digit by pressing ENTER.
e) Once the last digit is confirmed the scale return in weighing mode.

6.3 CALIBRATION WITH NO ZERO ACQUISITION AND NO CAPACITY SELECTION

PREMISE: Set the gravity of calibration and use zones (see section 6.2). Keep pressed the CALI. key for about 4 seconds.
Calibrate after the weight window displays "CAL.". There are two ways to calibrate:
1. Full range load: Put enough weight on the platform according to the full scale. Press ENTER to make sure.
2. Random load:
a. Press SET.
b. Input the weight you want to load by increasing the digit with the
successive digit by pressing ENTER.
c. Then load the corresponding weight. Press ENTER to make sure. Press MODE to exit from the calibration state in
the process of calibrating.

6.4 CALIBRATION WITH ZERO ACQUISITION AND CAPACITY SELECTION

PREMISE: Set the gravity of calibration and use zones (see section 6.2).
1. Turn off/on the scale.
2. Press both the ENTER and CALI. keys during the turning on.
3. By pressing the
4. Unload the scale and press ENTER.
5. Load a weight equal to the half capacity and press ENTER.
6. Load a weight equal to the capacity and press ENTER.

6.5 WEIGHING FILTER

1. Press both the MODE and SET keys at start-up, during self-checking.
2. By pressing the
its intervention.
3. Confirm with ENTER.
key one selects the available scale capacities: 1.5, 3, 6, 15, 30, 60. Confirm with ENTER.
key one selects the available filter: SPd 0, SPd 1, SPd 2. The higher the filter value, and greater is
to select 3000, 6000 or 7500; in the
10
key, and by confirming it
key, and by confirming it to set
key, and by confirming it to set the