Operation
1.
Set the selector according to the Settings Table.
2.
Refer to "Typical Measurements" for specific measurement instructions.
3.
Test the unit on a known functioning circuit or component.
• If the unit does not function as expected on a known functioning circuit, replace the
battery (or batteries).
• If the unit still does not function as expected, send the unit to Greenlee for repair. Refer to
the instructions under the Warranty.
4.
Take the reading from the circuit or component to be tested.
Settings Table
To measure
this value ...
Continuity*
Resistance
DC Voltage
AC Voltage
AC Current (1000 A max.)
Frequency
* Tone sounds if the measured resistance is less than approximately 30 Ω.
Electric shock hazard:
Contact with live circuits can result in severe injury or death.
set the selector
to this symbol ...
Hz
CM-1300 • CM-1350
connect the
red lead to ...
V-Ω
V-Ω
V-Ω
V-Ω
Remove lead
Remove lead
and connect the
black lead to ...
COM
COM
COM
COM
Remove lead
Remove lead
9