Adjusting the desired volume
Dispense volumes for the AutoRep M are
shown in Table 1 below. The table is a matrix
of all sizes of Encode tips and all stroke
settings.
The dispensed volume is obtained by multip-
lying the smallest volume unit (printed on the
Encode tip) by the stroke setting.
Example: 5 mL Encode tip
Smallest volume unit:
1 = 100 µL
Stroke setting 2:
100 µL x 2 = 200 µL /step
Dispensed volume µL
Table 1
Stroke setting
1
Number of steps
49
Encode tip size
Dispensed volume µL
0.5 mL
10
1.25 mL
25
2.5 mL
50
5 mL
100
12.5 mL
250
25 mL
500 1000 1500 2000 2500
50 mL
1000 2000 3000 4000 5000
2
3
4
5
24
15
11
9
20
30
40
50
50
75
100
125
100
150
200
250
200
300
400
500
500
750 1000 1250
Ajustage du volume désiré
Les volumes de dosage pour l'AutoRep M
sont indiqués dans la table 1 ci-dessous.
La table est une énumération de toutes les
tailles de pointes Encode tips et de toutes les
5
4
courses ajustables du piston.
3
2
Le volume dosé est obtenu en multipliant
1
l'unité de volume la plus petite (imprimée sur
la pointe Encode tip) par la course ajustée.
Exemple: pointe Encode tip de 5 mL
Table 1
5 mL
^
1
= 100 µL
Unité de volume la plus petite:
1 = 100 µL
Course ajustée 2:
100 µL x 2 = 200 µL /opération
Volume dosé µL
Course ajustée
1
Nombre d'opérations
49
Taille pointe Encode tip
0,5 mL
10
1,25 mL
25
2,5 mL
50
5 mL
100
12,5 mL
250
25 mL
500 1000 1500 2000 2500
50 mL
1000 2000 3000 4000 5000
2
3
4
5
24
15
11
9
Volume dosé µL
20
30
40
50
50
75
100
125
100
150
200
250
200
300
400
500
500
750 1000 1250