Cable, extension rod,
case
Snap Head
The snap head fixing system renders it very simple to fit a new impeller to the
flow probe: this saves costly non-productive time and in the event of damage to
the snap-head enables you to have the probe back in service in a matter of
seconds by simply changing the snap-head.
Checking / factory inspections
Depending on the application and the loads on the instrument, we consider it
highly advisable to have the instrument, respectively the probe annually checked
or inspected / repaired at our factory.
30
3) Leave the top part of the impeller to dry for ca. 30
minutes.
Only clean the head of the impeller!
Probes are on no account to be fully immersed
in the solution.
The following agents:
- Soapy water
- EDP plastic cleaner
- Window cleaning fluid (without ammonia)
- and for any stubborn deposits,
pure benzene
These solutions are only to be applied to the parts
with a fluff-free cloth.
Only carry out surface cleaning! The parts are
never to be immersed in the solution!