Hygiene and maintenance
W&H recommends sterilization according to EN 13060, EN 285 or ANSI/AAMI ST55.
> Read the notes, follow the instructions and heed the warnings provided by the manufacturers of steam
sterilizers.
> The program selected must be suitable for the motor.
Recommended sterilization procedures
> "Dynamic-air-removal prevacuum cycle" (type B) / "Steam-flush pressure-pulse cycle" (type S)*/**
134 C (273 F) for at least 3 minutes, 132 C (270 F) for at least 4 minutes
> "Gravity-displacement cycle" (type N)**
121 C (250 F) for at least 30 minutes
> Maximum sterilization temperature 135°C (275°F)