DKL L2-S600 Notice D'utilisation page 100

Publicité

Evidence of the medical device's basic suitability for effective sterilization was provided by an independent
test laboratory using the LISA 517 B17L* steam sterilizer (W&H Sterilization S.r.l., Brusaporto (BG)), the
Systec VE-150* steam sterilizer (Systec) and the CertoClav MultiControl MC2-S09S273** steam sterilizer
(CertoClav GmbH, Traun).
"Dynamic-air-removal prevacuum cycle" (type B):
"Steam-flush pressure-pulse cycle" (type S):
"Gravity-displacement cycle" (type N):
Drying times:
"Dynamic-air-removal prevacuum cycle" (type B):
"Steam-flush pressure-pulse cycle" (type S):
"Gravity-displacement cycle" (type N):
* EN 13060, EN 285, ISO 17665
** ANSI/AAMI ST55 , ANSI/AAMI ST79
134°C (273°F) – 3 minutes*, 132°C (270°F) – 4 minutes*/**
134°C (273°F) – 3 minutes*, 132°C (270°F) – 4 minutes*/**
121°C (250°F) – 30 minutes**
132°C (270°F) – 30 minutes**
132°C (270°F) – 30 minutes**
121°C (250°F) – 30 minutes**
31