INSTRUCTIONS FOR CLEANING AND STERILIZATION OF MINILIGHT, MINIMATE,
MINIBRIGHT SYRINGES IN ACCORDANCE WITH UNI EN ISO 17665 REQUIREMENTS
Warning
Preparation
Manual cleansing
Automatic cleansing
Manual disinfection
Automatic disinfection
Manual drying
Automatic drying
Maintenance and checking
Packaging
Sterilization
Preservation
• Inappropriate sterilization is dangerous for patients and operators.
• Do not perform spray sterilization. Do not place in any disinfectant liquids.
• Do not perform cold chemical or hot air sterilization.
• The appointed staff must be specialized and trained.
manufacturer indicated on the label.
• Do not use chlorine-based solutions.
• When sterilizing more than one piece in one autoclave do not exceed its
maximum load capacity.
Remove dirt from the surface using a disposable paper tissue.
the internal channels.
Press the button locateci at the bottom of the sleeve and slide the stainless
steel sleeve off the syringe body.
Unscrew the ferrule and remove the tip.
Wipe with a disposable paper tissue and with the aid of drinking water to
remove any impurities and dirt
Not available
Perform disinfection only with a disposable tissue and with compatible
disinfectants (in accordance with the instructions included in the product label
and technical data sheet). Recommended disinfectants:
• lncidin liquid
• FD 322 Durr
• Mikrozid AF Liquid
Not available
Dry with disposable paper tissue.
Dry with dry, clean and uncontaminated compressed air both internally and
externally until completely dry.
Do not dry with hot air.
Not available
No particular maintenance is required.
There is no objective term limiting the usage life of the sleeve: check to see if
there are if any damages or signs of wear and tear, replace the part if necessary.
The sleeve and tip are autoclavable.
Sterilize using EN 13060 ISO 17665-1 compliant class-B steam autoclave.
pressure, 4 minutes wait.
Never exceed 135° C.
The autoclave must be validated.
No particular requirements other than keeping them in their sealed and
sterilized package. Keep them in an appropriate environment and out of direct
sunlight and in a dry place, which should have low bioburden where possible.
13